  펄스 UV (PULSE UV) 살균시스템

 

 

 

 

 

 

 

 

펄스UV 살균시스템 사양 (PULSE UV STERILIZATION)

SteriPulse-XL Systems

 

 

SteriPulse-XL® RS-3000 Sterilization Systems

 

 img93.gif

 Model RS-3000C Sterilization Chamber

 

 

APPLICATIONS

 

• Sterilizing or decontaminating food, medical   devices, packaging materials, blood or   pharmaceutical  products

 

• Validate operating parameters for inline   processing

 

 

FEATURES

 

• R&D Benchtop (Model RS-3000C) and inline   processing systems available Models RS-3000B   and RS-3000M)

 

• Pulsed UV lamp delivers 505 joules/pulse

   (1.27 joules/cm²)

 

• Short pulses (360μs width) prevent excessive   heat buildup and accomplish sterilization without   overheating the product

 

• Front-panel operator control (Models RS-3000B   and RS-3000C) or PLC control (Model RS-3000M)

 

• No toxic materials, such as mercury or   microwaves are involved

 

 

GENERAL DESCRIPTION

 

The SteriPulse-XL Systems are high energy, pulsed UV light sterilization systems that have been shown to completely eradicate microorganisms (1) and provide a higher rate of sterilization than continuous mercury UV exposure. A revolutionary xenon lamp at the heart of the system delivers high energy pulses of UV light that penetrate deeply to achieve unprecedented degrees of decontamination. The pulsed UV light process is environmentally benign since it does not create or use volatile organic compounds (VOCs) or create suspended airborne particulates.

 

 

PULSED UV LAMP

 

Rapid, high intensity, broad spectrum UV light emitted by the clear fused quartz lamp causes formation of Pyrimidone dimers in DNA, resulting in genetic damage to cells and their ultimate destruction. Types of damage induced by pulsed light are: (a) photolysis; (b) loss of colony-forming ability (death); (c) inability to support phage growth (enzyme inactivation) and (d) destruction of nucleic acid.

 

 

STERILIZATION CHAMBER

 

The SteriPulse-XL Sterilization Chamber, with removable Lamp Housing, is detached from the Controller. Ozone is evacuated to ensure EPA ozone level compliance in the workplace. Forced air evacuation is in the range of 1-4 volumes per minute to ensure no heat buildup within the chamber during sterilization. The Chamber does not evacuate air borne pathogens nor introduce air borne pathogens into the Chamber while the system is in the ON cycle. Microbe filters, ozone resistant, are used at both the inlet and outlets of the ventilation path to ensure containment of air borne pathogens. The Chamber has an interlocking door for lockout during the sterilization ON cycle. The door interlock switch is connected with the safety interlock switches in the Controller. The Chamber and insert are made of stainless steel construction for ease of sterilization and disinfection. This construction also insures the Chamber is able to withstand UV light and heat conditions, experienced under normal operating conditions.

 

 

 

 

SPECIFICATION

All specifications are typical unless otherwise noted (TAMBIENT @ +25C, VINPUT = 208 VRMS, blower operating)

System Components

  Controller

Sterilization lamp power and front panel operator controls

  Lamp Housing¹

Lamp Housing with 16" Linear UV lamp

  Sterilization Chamber

Lamp Housing mounted on top; access door on front

  LiteMark-XL 1000

Lamp intensity monitor; mounted on Lamp Housing

  Blower Kit²

Provides filtered air to the Lamp Housing to cool lamp

  Interconnecting Cables

10-ft Lamp Control cable attached to Lamp Housing

 

10-ft High Voltage cable attached to Lamp Housing

 

 

 

Controller - Supplied with Models RS-3000B and RS-3000C

  Front Panel Controls

 

    Timer Power

ON/OFF

    High Voltage

ON/OFF

    Continuous mode

ON/OFF

    Pulse Mode Select

Timed or Continuous

    Programmable Timer

1 to 999 seconds in 1 sec intervals

    Timed Start

Pushbutton

 

 

  Mains Line Voltage

200-240 Vrms (±10%), (50/60 Hz), single phase

  Line Current

20 ampere @ 50 Hz, 18 ampere @ 60 Hz, maximum

  Mains Line Power

2500 W, maximum

  Mains Power Cord

8-ft (2.4 meters)

  Warm-up time

1 minute

  Outline Dimensions

8.0" x 18.5"x 27.0" (203 x 470 x 686 mm) (H x W x D)

  Weight

90 pounds (40.8 kg)

 

 

 

 

Lamp Housing

 

  Pulsed UV lamp

16-inch linear clear fused quartz; non-toxic; mercury free

  Radiant Energy

1.27 Joules/cm2 @ 0.76" (19.3 mm) from window face

  Pulse UV Lamp Footprint

16" x 1" (406.4 x 25.4 mm)

  Reflector Type

Ellipsoid

  Mounting Position

May be mounted in any position; cooling must be provided

  Air Cooling

Minimum 300 cfm of filtered air at 2.0" water gauge (free flow)

  Outline Dimensions

7.5" x 30.0" x 7.0" (190.5 x 762 x 178.8 mm) (H x W x D)

  Weight

36 pounds (16.35 kg)

 

 

 

 

Sterilization Chamber - Supplied with Model RS-3000C

  Access Door

Hinged; 18" x 6" (457.2 x 152.4 mm)

  Material

Metal grade stainless steel with clear passivate overcoat

  Disinfecting Methods

Note 4

  Door Interlock

High Voltage disabled at controller when chamber door open

  Chamber interlock  cable

2-ft (0.6 meters)

  Mains Power Cord

7-ftt (2.1 meters)

  Outline Dimensions

11.61" x 30.5" x 16.0" (294.9 x 774.7 x 406.4 mm) (H x W x D)

  Voltage

  (includes  internal fan)

200-240 Vrms (±10%), (50/60 Hz), single phase, 3 amperes

  Weight

57 pounds (25.9 kg)

 

 

 

 

System Environmental

  Temperature, ambient

 

    Rated Performance

0 to +40 oC (+32 to +104 oF)

    Storage

-40 to +85 oC

  Relative Humidity

10 to 80% @ +40 oC noncondensing

 

 

 

 

 

 

Notes

1. Supplied as a stand alone unit (models RS-3000B/M) or mounted on Sterilization Chamber

   (model RS-3000C)

2. Blower kit includes blower, blower filter, metallic ducting, duct clamps and mains power cord

3. Pulse rate is factory set and cannot be changed by user

4. The sterilization chamber and tray can withstand sterilizing or disinfecting methods such as autoclaving

   (276°F steam at 30 psi for 30 minutes), glutaraldehyde (Cydex) and/or 7% chlorine bleach disinfectant

   wash Specifications subject to change without notification.

 

 

 

MODEL CONFIGURATIONS

 

The SteriPulse-XL is available in three application specific configurations.

Model RS-3000B is a basic sterilization system offering a Controller unit and a standalone 16" Lamp Housing with Clear Fused Quartz (CFQ) lamp. The Controller provides all power to the Lamp Housing as well as complete operator control of the sterilization process using front-panel controls.

 

img95.gif

Model RS-3000B Controller, 16" Linear Lamp Housing and LiteMark-XL mounted on housing

 

Model RS-3000C is a complete benchtop sterilization system with Controller unit and separate Sterilization Chamber containing a 16" Lamp Housing and Clear Fused Quartz (CFQ) lamp. The Controller provides all power to the Lamp Housing as well as complete operator control of the sterilization process using front-panel controls. Safety interlocks are provided to protect the user from exposure from the lamp when the chamber door is open.

 

img99.gifModel RS-3000M is a modular system designed for ease of integration into large scale sterilization systems. Units include a Controller, High Voltage Power Supply Module, and a standalone 16" Lamp Housing with CFQ lamp. System control is via a remote PLC.

 

img97.gif

Model RS-3000M Controller, Power Supply, 16" Linear Lamp Housing and LiteMark-XL mounted on lamp housing

 

LAMP HOUSING COOLING

 

The 16" Linear Lamp Housing has been designed for continuous forced air cooling to insure the lamp temperature is maintained during operation. Air volume of 300 cubic feet per minute is required. Light blocking air exhaust filters are provided on the housing. To provide adequate cooling for the Lamp Housing and lamp, a blower kit is available.  

 

 

This kit includes a blower, mains power cord, air filter, flexible aluminum ducting and hose clamps.

 

 

INLINE STERILIZATION SYSTEMS

 

The SteriPulse-XL RS-3000B and RS-3000M systems can be applied in inline production applications, offering rapid sterilization. Model RS-3000M systems can be configured with multiple Lamp Housings to effectively increase the linear radiation length. The small bore lamp and elliptical reflector provide high peak UV irradiance for deep penetration and high speed sterilization, as well as maximum energy for dose-dependent sterilization. The ability to implement remote Programmable Logic Control (PLC) for timed sterilization (stop and sterilize), permits ease of integrating RS-3000M systems into small and large scale manufacturing systems

 

 

LITEMARK-XL LIGHT MONITOR

 

A useful accessory item that can be supplied with the SteriPulse-XL systems is the LiteMark-XL Light Monitor. The LiteMark-XL Monitoring System is a photo-electric detector module which is factory supplied mounted on a Lamp Housing or Sterilization Chamber to enable the operator to monitor, on a real-time basis, the performance of a Pulse UV Lamp system during its operating life. Such a capability allows the Pulse UV Lamp to be changed before the output falls below a predetermined safe minimum. It does this by sensing the light intensity from each flash which is scattered sideways in the Lamp Housing window and relating it to the side-scattered intensity produced by the same Pulse UV Lamp when new. The side-scattered intensity produced by the new Pulse UV Lamp is designated as the “100% level”, and the intensity at any later time is compared to that 100% level to produce a percentage value slowly declining from 100% as the lamp continues in use, indicative of the status of the Pulse UV Lamp at any given time. A correlation chart must be used in conjunction with the LiteMark-XL data to obtain the reduced exposure percentage.

 

img98.gif

LiteMark-XL Mounted on Lamp Housing

 

MEASUREMENT OF TARGET INTENSITY

 

Since the LiteMark-XL detector does not look directly at the Pulse UV Lamp and is not located at the actual target surface, it does not provide an absolute calibration in joules/cm² of the energy striking that target surface. It does, however, provide an indirect and real-time means of tracking the target surface exposure at any time if the new-lamp exposure value (the “100% level”) has been measured at the target. The correspondence between the actual target surface exposure at any time and the reduced value derived from the LiteMark-XL relative value and the correlation chart will remain in close agreement during the entire effective operating life of the Pulse UV Lamp.

The starting exposure (the 100% level) at the target surface must be measured with an exposure (joules/cm2) measuring instrument having a current calibration certification. The instrument calibration certification should cover the spectral range from 200 nm to 3 microns, the nominal transmittance range of the quartz components in the light path, except in cases where narrow-band exposure data is needed. It should be noted that this real time monitoring procedure has taken place without having to interrupt operation to take an irradiance measurement with a radiometer.

 

 

DATA LOGGER SYSTEM

 

The incorporation of the Data Logger System into the LiteMark-XL Light Monitoring System enables the evaluation process to be carried on continuously without the need for operator calculations The Data Logger is a computerized data collection and analysis system which captures, records, processes, stores and prints radiant energy information from any Xenon Corporation Lamp Housing or Sterilization Chamber fitted with a LiteMark-XL Light Monitoring system. The Data Logger interfaces to a personal computer through an Analog-to-Digital Converter (ADC) which is built into the output connector of the Data Logger cable, plugging directly into the computer printer port. The system operator can observe the Pulse UV Lamp intensity display on his computer monitor as either tabulated data or as a graph. Additionally the operator can input the data into an Excel-style spreadsheet for further processing off-line.

 

 

    

 

저온UV경화, 플래시UV, 펄스UV, PULSE UV, XENON FLASH UV LAMP, XENON, DVD BONDING, WAFER BONDING, 제논, 웨이퍼, 표면개질, UV세정, UV CLEANING, 엑시머, EXCIMER 세정, 필름세정, 무전극UV, MICROWAVE UV, ELECTRODLESS UV, FUSION UV,COOL UV, FAST CURING,SPOT UV 자외선 조사기, UV SPOT CURE, 의료기구, 튜브접착, 자동화라인, 고속건조, 자동본딩,살균, 위생, 식품처리, DNA, RNA, 스크린인쇄, 프린팅, 코팅, 점착, screen printing, coating, adhesive, lcd glass, touch screen, film, pmma, UV Curing, Ultraviolet Curing Systems, UV Curing, UV Curing Equipment for Medical Assembly, Medical Assembly, Tubing bonding, UV,UV Rotary Curing, UV Curing Technology, Ultraviolet Curing Technology, Automated Process Lines, Disk Drive HGA Bonding, High Speed Automated Curing, High Speed Automated Assembly, Ink Jet Cartridge Manufacturing, Automated Bonding

 

 